14
Default Settings
Parr Instrument Company
48
Calorimeter Operation
Stirrer OFF
Operating Controls
Method of Operation Solution
LCD Backlight Timeout(s) 1200 S
LCD Backlight Intensity 70%
Print Error Messages ON
Language English
Program Info and Control
Settings Protect OFF
Data Entry Controls
Auto Sample ID Controls ON
Auto Sample ID Controls
Automatic Sample ID ON
Automatic Sample ID Number 1
Automatic Sample ID Increment 1
Communication Controls
Printer
Printer Type Parr 1758
Printer Destination Local (USB)
Balance Port
Balance Port Device /dev/tty/USB0
Number of Data Bits 8
Parity None
Number of Stop Bits 1
Handshaking None
Baud Rate 9600
Data Characters from Balance 8
Data Precision 4
Transfer Time Out 10
Diagnostics
Calorimetry Parameters
Tolerance (L) Parameters
Preperiod Tolerance (L1) 0.002
Misfi re Tolerance (L2) 0.05
Dynamic Post period Tolerance (L3) 0.002
Eqlbrm Post period Tolerance (L4) 0.002
Preperiod Stability (L5) 2
Preperiod Blackout (L6) 36
Derivative Wghtng Cutoff (L7) 0.02
Preperiod Timeout (L8) 42
Postperiod Timeout (L9) 60
Preperiod Tolerance Jck Calc’d (L10) 0.5
Correction (K) Parameters
K1 0.5
K2 0.00080
K3 1.0
K4 0.0
K5 0.0
Blackout (B) Parameters
Misfi re Blackout (B2) 72
Derivative Blackout (B3) 0.5
Dynamic Blackout (B4) 6
Equilibrium Blackout (B5) 18
Dynamic Derivative Blackout (B6) 0.02
Dynamic Time Blackout (B7) 20
Data Logger
Data Logger OFF
Interval in Seconds 10
Data Log Destination Log File and Printer
Data Log Format Text
Reporting Controls
Report Width 40
Automatic Reporting ON
Auto Report Destination Printer
Individual Printed Reports OFF
CHAPTER 14
Default Settings
The following are recommended operation settings for the 6755 Solution Calorimeter.