Communications Interfaces
6100
D
www.parrinst.com
57
Generic Interface
The data field should consist
of 9 numeric characters (0
through 9, +, - and space)
terminated with a carriage
return (CR). Leading zeros may
be blanked as spaces and are counted. Non-numeric
characters are ignored and will reset the input buffer
if the data field has not been filled. Any characters
received after filling the data field and before the
carriage return are ignored.
Table D-1
6100 Data File Naming Convention
Test data files are named with the following convention.
Test Type Filename
Preliminary Standardization <ID>.std.plim.csv
Final Standardization <ID>.std.finl.csv
Preliminary Determination <ID>.det.plim.csv
Final Determination <ID>.det.finl.csv
Pre-weigh <ID>.---.pwgh.csv
Table D-2
6100 Calorimeter Run Data Template
Field Description
SampleID char[16]
Timestamp MM/DD/YY HH:mm:ss
Mode 0 = determination, 1 = standardization
Method 0 = equilibrium, 1 = dynamic
State 0 = preweigh, 1 = preliminary, 2 = final
Units 0 = MJ/kg, 1 = Btu/lb, 2 = cal/g,
3 = J/kg, 4 = other
UnitMultIfOther unit multiplier in effect at time of
report
BombID [1,4]
BombEE bomb energy equivalent
SampleWt sample weight
SpikeWt spike weight
Fuse fuse value
FuseFinal fuse value is final
Acid acid value
AcidFinal acid value is final
Sulfur sulfur value
SulfurFinal sulfur value is final
Hydrogen hydrogen value (net calc option)
HydrogenFinal
hydrogen value is final (net calc option)
MAD moisture as determined value (dry
calc option)
MAD Final moisture as determined is final
JacketTemp jacket temperature
InitTemp initial temperature
DeltaT temperature rise
HOC gross heat of combustion
NetHOC dry net HOC (net calc options enabled)
DryHOC dry gross HOC
(if dry calc option enabled)
DryNetHOC dry net HOC (if both dry and net calc
options enabled)
Oxygen oxygen value (net calc option)
Oxygen Final oxygen value is nal
Nitrogen nitrogen value (net calc option)
Nitrogen Final nitrogen value is final
MAR moisture as received (dry calc option)
MAR Final moisture as received value is final
Dry Net HOC_AR
Dry net HOC as received value (if both
dry and net calc option enabled)
Bomb Name bomb name assigned to bomb ID
Field Length
data 9
CR 1